BioProject
오믹스 분석을 통한 환경보건취약지역 주민의 환경성질환 원인물질 추적 기술개발
-
AccessionKAP241552
-
Submission date2025-07-17
|
Technique type
|
Mass spectrometry (MS) |
|---|---|
|
Assay Type
|
Chromatography (MS-based) |
|
Species
|
Homo sapiens |
|
샘플 유형
|
Human |
|
추출 프로토콜
|
To extract adducts from urine, 1 mL of each sample was transferred into a 1.5 mL microcentrifuge tube. Prior to sample loading, the Sep-Pak C18 column(waters seppak vac 1cc(100mg)C18 cartridges) was conditioned by passing 1 mL each of methanol and water. The sample was then loaded onto the column, followed by purification and extraction using 1 mL each of water and a methanol/ethyl acetate mixture. The eluted solution was evaporated and concentrated at 60°C for approximately 20 minutes using a nitrogen concentrator. After evaporation and concentration, the sample was reconstituted with the solvent used in the pretreatment process, filtered through a syringe filter, and transferred to an autosampler vial for analysis.
|
|
크로마토그래피 프로토콜
|
LC separations were performed on a Hypersil GOLD VANQUISH column (100 × 2.1 mm, particle size 1.9 μm, Thermo fisher). The column temperature and flow rate were set at 30°C and 0.3 ml/min, respectively. The mobile phase used contained 0.1% formic acid in ammonium formate buffer (0.1mM) (A) and acetonitrile (100, v/v) (B). The linear gradient was as follows. 1.0% B for 1 min, 1.0-10% B for 1 min, 10-100% B for 8 min, 100% B for 2 min, 100-1.0% B for 8 min. The injection volume of the sample was adjusted according to the positive ionization polar mode and It was 5 μl using partial loop mode for both negative ionization polar modes. Quality control (QC) samples pooling equal aliquots of samples were measured regularly throughout the run to ensure data reproducibility.
|
|
질량분석 프로토콜
|
To obtain MS spectral data, LC-ESI-MS/MS analysis was performed on a TSQ Altis Plus (Thermo scientific) coupled with a UPLC system (Thermo scientific).
|